| Model | Calsys -100/40 / Autocal |
|---|---|
| Temp Range | -100°C ~ 40°C |
| Stability (30 Min) | ±0.04 °C or Better |
| Radial Uniformity | ±0.05°C or better |
| Immersion Depth** | 150 mm |
| Method of Control | Self tunned PID controller |
| Cooling Time | 70 Min ( Ambient 25°C to -100°C) |
| Resolution | 0.1 °C (0.01 Optional) |
| Computer Interface | RS - 232 |
| Input (CALsys -100/40 Autocal) | Four Channels (One master and three test sensors), high-quality LEMO connector suitable for both T/C (J, K, N, T, R, S, B type) and Rtd. |
| Software (CALsys -100/40 Autocal) | The calibrator should be provided with software for data recording (Manual Mode) and Test Certificate generation in Auto Mode |
| Data logging (CALsys -100/40 Autocal) | Data logging facility with logged data export to computer through LAN port (optional USB) |